Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Data Acquisition Time' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Data Acquisition Time' found in 1 term [] and 2 definitions [], (+ 19 Boolean[] results
previous     16 - 20 (of 22)     next
Result Pages : [1]  [2 3 4 5]
Searchterm 'Data Acquisition Time' was also found in the following service: 
spacer
News  (3)  
 
Partial Echo
 
(PE) The partial echo technique (also called fractional echo) is used to shorten the minimum echo time. By the acquisition of only a part of k-space data this technique benefits (like all partial Fourier techniques) from the complex conjugate symmetry between the k-space halves (this is called Hermitian symmetry).
The dephasing gradient in the frequency direction is reduced, and the duration of the readout gradient and the data acquisition window are shortened. Partial echo gives a better SNR at a given TE when a smaller FOV or thinner slices are selected, allows a longer sampling time, and a larger water fat shift (WFS, see also bandwidth) due to a lower gradient amplitude. The resolution is not affected. This is often used in gradient echo sequences (e.g. FLASH, Contrast Enhanced Magnetic Resonance Angiography) to reduce the echo time and yields a lower gradient moment. The disadvantage of using a partial echo can be a lower SNR, although this may be partly offset by the reduced echo time.
Also called Fractional Echo, Read Conjugate Symmetry, Single Side View.

See also Partial Fourier Technique and acronyms for 'partial echo' from different manufacturers.
spacer
 
Further Reading:
  Basics:
Method and apparatus for subterranean formation flow imaging
   by www.google.com    
MRI Resources 
Spectroscopy pool - General - Libraries - Calculation - Education - Equipment
 
Point Resolved SpectroscopyInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PRESS) Point resolved spectroscopy is a multi echo single shot technique to obtain spectral data. PRESS is a 90°-180°-180° (slice selective pulses) sequence. The 90° radio frequency pulse rotates the spins in the yx-plane, followed by the first 180° pulse (spin rotation in the xz-plane) and the second 180° pulse (spin rotation in the xy-plane), which gives the signal.
With the long echo times used in PRESS, there is a better visualization of metabolites with longer relaxation times. Many of the metabolites depicted by stimulated echo technique are not seen on point resolved spectroscopy, but PRESS is less susceptible to motion, diffusion, and quantum effects and has a better SNR than stimulated echo acquisition mode (STEAM).
spacer

• View the DATABASE results for 'Point Resolved Spectroscopy' (3).Open this link in a new window

 
Further Reading:
  Basics:
The Basics of MRI
   by www.cis.rit.edu    
  News & More:
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
MRI Resources 
Pediatric and Fetal MRI - MRI Physics - Absorption and Emission - Contrast Agents - MRI Reimbursement - Equipment
 
Rapid Excitation Magnetic Resonance ImagingInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(RE MRI) There are several approaches to speeding up the MRI data acquisition process by repeating the excitation by RF pulses in times short compared to T1, typically using small flip angles and gradient echo refocusing. When TR is also on the order of or shorter than T2, the repeated RF pulses will tend to refocus transverse magnetization remaining from prior excitations, setting up a condition of steady state free precession, and a dependence of signal strength (and image contrast) on both T1 and T2.
This can be modified in various ways, particularly:
1) to spoil the tendency to build up a steady state by reducing coherence between excitations, e.g. by variation of the phase or timing of consecutive RF pulses or of the strength of spoiler gradient pulses, thus increasing the relative dependence of signal strength on T1 or
2) acquire the signal when it is refocusing immediately prior to the next RF pulse, thus increasing the relative dependence of signal strength on T2.

See also Ultrafast Gradient Echo Sequence.
spacer

• View the DATABASE results for 'Rapid Excitation Magnetic Resonance Imaging' (2).Open this link in a new window

Searchterm 'Data Acquisition Time' was also found in the following service: 
spacer
News  (3)  
 
Rectangular Field of View
 
(RFOV) A different field of view (the scanned region) in the frequency and phase encoding directions that means the data acquisition with fewer measurement lines. Because there are fewer rows than columns, a rectangular image is obtained. To reduce the FOV in phase encoding direction (foldover direction) saves scan time by decreasing signal but invariable spatial resolution.
Also called HFI or undersampling.
mri safety guidance
Image Guidance
If the scanned object is oval, e.g. head or abdomen, a rectangular FOV is an easy to use scan parameter to reduce the scan time without loss of resolution.
spacer

• View the DATABASE results for 'Rectangular Field of View' (2).Open this link in a new window

MRI Resources 
Hospitals - MRCP - Open Directory Project - Shoulder MRI - - Jobs pool
 
Scan TimeForum -
related threads
 
(SCT) The total scan time is the time required to collect all data needed to generate the programmed images. The scan time is related to the used pulse sequence and dependent on the assemble of parameters like e.g., repetition time (TR), Matrix, number of signal averages (NSA), TSE- or EPI factor and flip angle.
For example, the total scan time for a standard spin echo or gradient echo sequence is number of repetitions x the scan time per repetition (means the product of repetition time (TR), number of phase encoding steps, and NSA).

See also Number of Excitations, Turbo Spin Echo Turbo Factor, Echo Planar Imaging Factor, Flip Angle and Image Acquisition Time.

See also acronyms for 'scan time parameters' from different manufacturers.
spacer

• View the DATABASE results for 'Scan Time' (47).Open this link in a new window


• View the NEWS results for 'Scan Time' (10).Open this link in a new window.
 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
  News & More:
For MRI, time is of the essence A new generation of contrast agents could make for faster and more accurate imaging
Tuesday, 28 June 2011   by scienceline.org    
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Homepages - Implant and Prosthesis pool - Bioinformatics - Breast MRI - Safety pool - Nerve Stimulator
 
previous      16 - 20 (of 22)     next
Result Pages : [1]  [2 3 4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 2 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]